ES3001 Ground Resistance Soil Resistivity Tester

ES3001 Ground Resistance Soil Resistivity Tester

  • voltage range :0-100V
  • Soil resistivity range:0.00-9999kΩ
  • Earth resistance range:0.0-30000Ω
  • Measurement mothod : Precise four-wire、 three-wire measurement, simple two-wire measurement
  • USB interface: With USB interface, software monitoring, storage data can be uploaded to the computer, save and print

Technical Parameter

FunctionTwo three four-wire measure grounding resistance, soil resistivity;Ground voltage, AC voltage measurement
Ambient temperature and humidity23℃±5℃,below 75%rh
Power8.4V lithium battery (built-in)
Interference voltage<20V (should be avoided)
Interference current<2A (should be avoided)
Measure R electrode spacinga>5d
Measured electrode spacinga>20h
Auxiliary ground resistanceReference condition <100Ω, working condition <5kΩ
RangeGrounding resistance: 0.00Ω~30.00kΩ
Soil resistivity: 0.00Ωm~9999kΩm
Grounding voltage: 0.00V~100.0V
measurement modePrecise four-wire、 three-wire measurement, simple two-wire measurement
Measurement methodsGrounding resistance: rated current change pole methodSoil resistivity: four-pole methodGround Voltage: Average Rectification(between S-ES interface)
Test frequency128Hz
Short circuit test current> 20mA (Sine Wave)
Open circuit test voltageAC 28V max
Electrode spacing rangeCan be set 1m ~ 100m
Change gearfully automatic shifting 0.00Ω ~ 30.00kΩ
fully automatic shifting 0.00Ωm~9000kΩm
BacklightControllable gray screen backlight, suitable for use in dim places
Display mode4-bit large LCD display, gray screen backlight
Measurement instructionsLED flashing indicator during measurement
LCD size111mm×68mm
LCD display field108mm×65mm
Instrument sizeL/W/H:277.2mm×227.5mm×153mm
Standard test line length4 strips:red 15m,black 15m, yellow 10m,green 10m each one
Simple test line2strips:yellow 1.6m,green 1.6m each one
Auxiliary Grounding rod4PCS:φ10mm×200mm
Measure timeGround voltage:about 3 times/sec;
grounding resistance、soil resistivity: about 7 seconds/time
Line voltageMeasurement below AC100V (ground voltage measurement function cannot be used to measure commercial power)
USB interfaceWith USB interface, software monitoring, storage data can be uploaded to the computer, save and print
Communication LineOne USB communication line, 1.5m long
Data storage500 groups, “MEM” storage indicates, flashing “FULL” symbol indicates that the memory is full
Data reviewData review function: “MR” symbol display
Overflow displayOver-range overflow function: “OL” symbol display
Interference testAutomatic identification of interference signals, “NOISE” symbol indication when the interference voltage is higher than 5V
Auxiliary grounding testWith auxiliary ground resistance test function, 0.00KΩ ~ 30kΩ (Rh max = 3kΩ +100R <50kΩ; Rs max = 3kΩ +100R <50kΩ)
Alarm functionAlarm when the measured value exceeds the alarm setting value
Battery voltageReal-time display of battery power, reminding timely charging when battery voltage is low
Automatic Shutdown“APO” Indicates, automatic Shutdown After 15 Minutes
Power consumptionStandby: 40mA Max(Backlight off)
Turn on backlight: about 43mA
measuring:75mA Max(Backlight off)
WeightInstrument: 2450g(including battery)
Test lines:1300g
Auxiliary grounding rod: 850g(4PCS)
Working temperature and humidity-10℃~40℃;below 80%rh
Storage temperature and humidity-20℃~60℃;below 70%rh
Overload protectionGrounding resistance:AC 280V/3 seconds between H-E and S-ES ports
Insulation resistance20MΩ以上(500V between circuit and housing)
Pressure resistanceAC 3700V/rms(between circuit and housing)
Electromagnetic propertiesIEC61326(EMC)
Suitable for safety regulationsIEC61010-1(CAT Ⅲ 300V、CAT IV 150V、pollution level 2);IEC61010-031;IEC61557-1(grounding resistance);IEC61557-5(soil resistivity);JJG 366-2004.