FR100RD Series Rogowski Coil Current Sensor (with integrator)
FR100RD Series Rogowski Coil Current Sensor (with integrator)
- Jaw size: Φ8mm
- Range:AC 0~10KA
- Resolution:AC 0.1A
- Response frequency:0.1Hz~10MHz
- Line voltage:Line test below AC 1000V
- Output cable length:2m

Technical Parameter
| Features | Detection of AC leakage current, large current, high harmonic current, complex waveform current, transient impulse current, phase, electrical energy, power, power factor, etc. |
| Detection method | Flexible CT: The output signal is the differential of the current to time. By integrating the output voltage signal, the input current can be truly restored |
| Coil diameter | φ8mm |
| Range | 0A~10kA |
| Resolution | 0.1A |
| Output | 0.3mV/A 3mV/A 30mV/A third gear (0-5V, 4-20mA and other outputs can be customized) |
| Power supply | 9V 6F22 battery |
| Low battery indication | Have |
| Power consumption | The voltage output is 6 mA, and an alkaline dry battery is expected to continuously supply power for 80 hours |
| Accuracy class | ±1.0%FS (50Hz/60Hz); 23℃±2℃, below 70%RH, the wire is in the center of the coil) |
| Phase error | ≤1°(50Hz/60Hz; 23℃±2℃) |
| Output Interface | BNC connector (optional audio plug, banana plug, bare wire) |
| Output cable length | 2m |
| Electric field disturbance | No hysteresis effect, no external electric field interference |
| Wire position | The tested lead is in the center of the coil, and the influence of position error is ≤±0.3%FS |
| Response frequency | 0.1Hz~10MHz |
| Line voltage | Line test below AC 1000V |
| Working temperature and humidity | -20℃~50℃; below 80%rh |
| Storage temperature and humidity | -10℃~60℃; below 70%rh |
| Dielectric strength | AC 2000V/rms (both ends of the coil) |
| Suitable for safety regulations | IEC1010-1, IEC1010-2-032, Pollution Degree 2, CAT IV 1000V |
| Random accessories | Sensor: 1 piece |